Course Instructions

Objectives

Introduction

X-ray Characteristics

Density and Contrast (Imaging Terms)

Effects of Radiation

Film Speed

Digital Radiography

Infection Control

Prescribing Radiographs

Who Gets Which Films?

Shadow-Casting Principles

Radiographic Surveys

Patient Management

Film Processing

Qualities of Excellent X-Rays

Common Errors

Mounting Films

Glossary

Word Processor Printable Test and Answer Sheet

Web Based Test

Common Errors

Operator errors in film placement and angulation of the tubehead often result in undiagnostic x-rays. X-rays that are undiagnostic are useless to the dentist and must be retaken. Every effort should be taken to minimize the following errors, because each retake exposes the patient to more radiation.

Film Placement

  • Correct film placement is critical for success with x-rays. If the correct technique is followed every time, positioning errors will be minimized.

  • In all premolar views, the distal of the cuspid is visible.

  •  All molar views should contain the third molar region even if the tooth is not present in the mouth.

  • When focusing on a specific tooth, it should be centered on the film.

  • The film must be placed high enough in the palate or low enough in the floor of the mouth to clearly show the apex of the tooth in question and 3 to 4 mm of bone.

  • Films should not be bent if possible, but if the patient is uncomfortable with the edge, try gently reshaping the edge and repositioning the film in the mouth. Asking the patient to "gently" close often decreases the discomfort especially in the floor of the mouth.

  • Before placing an x-ray film in a patient's mouth, check to make sure that it is not backwards. The lead foil will leave an artifact on the exposed x-ray and it will be confusing to mount.

  • Make sure that exposed films are not mixed up with unexposed ones.

Angulation Of The Tubehead

Errors of angulation of the tubehead are common. When using an instrument, make sure that the tubehead is aligned correctly, parallel with the indicator rod and aligned with the ring if the operator is using a Rinn aparatus. If not using a Rinn, the beam should be parallel with the bitewing tab, or at an angle that splits the difference between the angulation of the film and the angulation of the tooth.  Film positioning devices are helpful and, when used correctly, they generally produce acceptable results.  When the patient's anatomy alters the usefulness of the instrument, it is best to revert to the the split angle technique. For example, if the patient has a shallow palate, and the instrument will not allow the film to be parallel with the long axis of the tooth, this is an ideal time to compensate using a split angle technique to avoid foreshortening.

Overlapping is another common error of angulation. If the cone is not perpendicular to the film, the contacts will be overlapped. Some offices routinely use a size 3 film for a bitewing view that contains all the teeth from premolar to molar. Due to the curve of the arch, some area is bound to be overlapped. It is better to position two size 2 films in a premolar view and then a molar view so all contacts will be opened. 

Cone cutting is another common error.  It happens mostly because of positioning the cone too far to the distal (mesial cone cuts are the most frequent kind). The film will be cone cut when the tubehead is not covering the whole area of the film. The best way to avoid this is by looking at the film in the patient's mouth and aiming the cone head directly toward the film instead of guessing from extraoral landmarks. Ask the patient to "grin wide" like the joker in the batman series so you can see down the buccal corridor (the area between the buccal surfaces of the teeth and the buccal mucosa).  This makes it much easier to aim the cone.

If the patient moves, the film will be affected. Watch the patient as you expose the film to make sure they don't move. The patient should be instructed to hold very still and not swallow until he or she is told it is OK.


Correct film processing procedures were previously discussed in the section on film processing, but the following table summarizes common film processing errors, the results, and possible solutions.

 

Error

Result

Solution

Developer temperature too low

Films too light

Check and adjust temperature

Developing time not long enough

Films too light

Use a timer

Developer solution too old or diluted

Films too light

Yellow or brown film

Keep a schedule of chemical maintenance

Change solutions when this begins to happen consistently

Developer too warm

Flim too dark

Check and adjust temperature

Developing time too long

Film too dark

Foggy film

Use a timer

Light leak in processing

Film too dark

Foggy film

Check processor and darkroom

Film exposed to light before processing

Film too dark

Foggy film

Don’t open film until safety light is turned on and other lights are off

Check safety light for leaks

Films exposed to radiation after exposure

Foggy film

Take exposed films out of room when exposing other radiographs

Fixer too old or contaminated

Yellow or brown film

Green film

Check with checking film

Replenish and maintain on schedule

Incorrect rinsing between developer and fixer

Streaking

Follow proper protocol for rinsing between chemicals

Chemicals exhausted

Incorrect or insufficient washing

Streaking

Green film

Run checker film every day

Wash longer after fixer

Dirty rollers, fingerprints

Artifacts or Streaking

Clean rollers, handle films by edge

Films sticking together

Green Film

Refix and rinse

Bending films

Lines on film

Don’t bend films

Static electricity

Lightning marks on film

Humidify environment, use Static Guard

Water drops on film

White spots

Don’t put film in water after removing it from it’s wrapper

Developer splash on film

Black spot

Make sure countertop is dry after changing chemicals

 


 <== Previous page       Next page ==>